測氫儀電解池作過(guo)一至二百個(ge)樣品(pin)后,應重新(xin)處理(li)(li),否(fou)則將直接影響氫值(zhi)的分析結果,使(shi)氫值(zhi)偏低(di),處理(li)(li)方(fang)法如下:
(1) 卸下電解池;
(2) 清(qing)(qing)洗(xi)(xi)(xi)(xi)電(dian)解池:首先用(yong)(yong)(yong)自來水將(jiang)其(qi)沖洗(xi)(xi)(xi)(xi),然后用(yong)(yong)(yong)小軟(ruan)毛刷(shua)沾上洗(xi)(xi)(xi)(xi)衣(yi)粉(fen),慢(man)慢(man)向(xiang)左向(xiang)里旋動(dong)毛刷(shua),注意旋動(dong)到引線即(ji)可,再慢(man)慢(man)地向(xiang)右向(xiang)外旋動(dong)毛刷(shua)退(tui)出,用(yong)(yong)(yong)水沖洗(xi)(xi)(xi)(xi),看鉑絲是否(fou)洗(xi)(xi)(xi)(xi)亮,如發(fa)現有黑處(chu),可重(zhong)新(xin)(xin)用(yong)(yong)(yong)毛刷(shua)刷(shua)洗(xi)(xi)(xi)(xi)直到洗(xi)(xi)(xi)(xi)凈(jing),最后用(yong)(yong)(yong)蒸餾水沖洗(xi)(xi)(xi)(xi)干凈(jing),再用(yong)(yong)(yong)丙銅沖洗(xi)(xi)(xi)(xi)一下,完畢后,用(yong)(yong)(yong)電(dian)吹風(熱風檔) 吹干,接著用(yong)(yong)(yong)萬用(yong)(yong)(yong)表(biao)電(dian)阻擋(2K) 量(liang)電(dian)解池兩極,內阻應量(liang)無窮大(da),表(biao)明清(qing)(qing)洗(xi)(xi)(xi)(xi)完好(hao),否(fou)則重(zhong)新(xin)(xin)清(qing)(qing)洗(xi)(xi)(xi)(xi)。
(3) 配電(dian)解液:取10ml或15ml的小(xiao)量筒,按體積比先盛入3份磷(lin)酸(優級(ji)純),再盛入7份丙銅(tong),輕搖量筒,筒底(di)應(ying)有熱量散發。
(4) 涂(tu)(tu) 膜:將洗清好的(de)電解(jie)(jie)池(chi)(chi)粗端傾斜向(xiang)上,把配好的(de)電解(jie)(jie)液(ye)向(xiang)電解(jie)(jie)池(chi)(chi)內(nei)倒,第一(yi)次(ci)(ci)到至(zhi)電解(jie)(jie)池(chi)(chi)長度的(de)1/2 (若(ruo)電解(jie)(jie)液(ye)還向(xiang)下流(liu)(liu)(liu),應(ying)將其放平,用(yong)(yong)(yong)濾紙擦凈(jing)流(liu)(liu)(liu)過(guo)來的(de)電解(jie)(jie)液(ye)),然(ran)后用(yong)(yong)(yong)冷風吹(chui)干(gan),第二次(ci)(ci)倒至(zhi)電解(jie)(jie)池(chi)(chi)全部 (注意不(bu)要流(liu)(liu)(liu)過(guo)兩(liang)極) ,再用(yong)(yong)(yong)冷風吹(chui)干(gan),第三次(ci)(ci)倒入(ru)一(yi)半,用(yong)(yong)(yong)冷風吹(chui)干(gan),然(ran)后用(yong)(yong)(yong)萬用(yong)(yong)(yong)表(biao)(1Ω擋(dang))量兩(liang)極,如(ru)有(you)的(de)阻值(幾歐(ou)或幾十歐(ou)),則涂(tu)(tu)膜正常,否則重涂(tu)(tu)。
(5) 裝(zhuang)(zhuang)上(shang)涂(tu)膜好的電(dian)解(jie)池(chi),并將冷卻套裝(zhuang)(zhuang)上(shang),使(shi)電(dian)解(jie)池(chi)細端略微向上(shang)傾(qing)斜(xie)。
(6) 電(dian)解(jie):打(da)開氣路,流量計顯示80mL/min,開啟電(dian)源,按(an)下涂膜(mo)鍵,60分(fen)鐘后(hou)電(dian)解(jie)至終(zhong)(zhong)點(dian)(dian),可(ke)再按(an)涂膜(mo)鍵,繼續涂膜(mo),直到終(zhong)(zhong)點(dian)(dian)(電(dian)解(jie)流≤70mA),到達終(zhong)(zhong)點(dian)(dian)后(hou),可(ke)按(an)“月、日(ri)”鍵,分(fen)別按(an)入單數和雙數,再按(an)返回(hui)鍵使電(dian)解(jie)電(dian)流能快速降至終(zhong)(zhong)點(dian)(dian),如此反(fan)復多次,即可(ke)使用。至此電(dian)解(jie)池處(chu)理完畢,做正式樣(yang)前先分(fen)析兩個(ge)廢樣(yang)以平衡系統(tong)。
7、注意事項:
(1) 將盛有試樣的瓷(ci)舟放入石(shi)英托盤上后,要將錐管蓋擰緊,以防(fang)漏(lou)氣,影響(xiang)測試結果。
(2) 檢查系統的氣密性時(shi),堵(du)口時(shi)間不可過(guo)(guo)長,以(yi)免(mian)內部氣壓過(guo)(guo)大,而引起崩塞現象。
(3) 如需(xu)將試樣提前拉出時,只須按一(yi)下返回(hui)鍵即可(ke),若打(da)印結(jie)果需(xu)要(yao)多份(fen)時,可(ke)在第一(yi)次打(da)印完畢后(hou),按一(yi)下打(da)印鍵,便可(ke)再得到一(yi)份(fen)打(da)印結(jie)果。
(4) 打印結果格(ge)式: 年 月 日(ri)、 重量 、 空白、 水分、 顯示(shi)H值(zhi)、 總氫(qing)值(zhi)、 分析基(ji)氫(qing)值(zhi)、 干基(ji)氫(qing)值(zhi)、 序號 。
(5) 清洗電(dian)解池時,毛刷(shua)不要刷(shua)過(guo)兩極引(yin)線,以免使(shi)電(dian)解池損壞。
(6) 控制器供電電壓,若出現過波動(dong)時(shi),可能(neng)會干擾程序運行,當顯示器出(chu)現異(yi)常顯示時(shi),可先(xian)按清(qing)零(ling)鍵清(qing)除(chu),若清(qing)除(chu)不掉,應(ying)關(guan)閉電源(yuan)開關(guan),稍后再(zai)重新(xin)啟動(dong)。
(7)實驗完畢后(hou),仍(reng)須將送樣錐管擰緊,并將電(dian)解池未(wei)端封閉,以防止潮濕空氣進入。