測(ce)(ce)試(shi)燃(ran)(ran)(ran)燒(shao)易于(yu)飛濺試(shi)樣的(de)發熱(re)(re)(re)量(liang)時,要(yao)先用(yong)(yong)已知質(zhi)量(liang)的(de)擦(ca)(ca)鏡紙(zhi)(zhi)包緊再進(jin)行(xing)測(ce)(ce)試(shi),或(huo)(huo)先在(zai)(zai)壓(ya)(ya)餅機中壓(ya)(ya)餅并(bing)切(qie)成(cheng)的(de)2-4mm的(de)小(xiao)塊使用(yong)(yong)。不(bu)易燃(ran)(ran)(ran)燒(shao)完(wan)全的(de)試(shi)樣,可(ke)先在(zai)(zai)燃(ran)(ran)(ran)燒(shao)皿(min)(min)底(di)墊(dian)(dian)(dian)上(shang)一個石(shi)棉墊(dian)(dian)(dian),或(huo)(huo)用(yong)(yong)石(shi)棉絨(rong)做襯(chen)(chen)墊(dian)(dian)(dian)(先在(zai)(zai)皿(min)(min)底(di)鋪上(shang)一層(ceng)石(shi)棉絨(rong),然后以手(shou)壓(ya)(ya)實),石(shi)英燃(ran)(ran)(ran)燒(shao)皿(min)(min)不(bu)需任何襯(chen)(chen)墊(dian)(dian)(dian),如(ru)加襯(chen)(chen)墊(dian)(dian)(dian)仍燃(ran)(ran)(ran)燒(shao)不(bu)完(wan)全可(ke)提高充氧壓(ya)(ya)力至3.2Mpa或(huo)(huo)用(yong)(yong)已知質(zhi)量(liang)和熱(re)(re)(re)值的(de)擦(ca)(ca)鏡紙(zhi)(zhi)包裹稱好的(de)試(shi)樣并(bing)用(yong)(yong)手(shou)壓(ya)(ya)緊,然后放入燃(ran)(ran)(ran)燒(shao)皿(min)(min)中。熱(re)(re)(re)容(rong)量(liang)測(ce)(ce)定(ding)的(de)苯甲酸(suan)應(ying)預先在(zai)(zai)盛有濃(nong)硫酸(suan)的(de)干(gan)燥器中干(gan)燥3天或(huo)(huo)在(zai)(zai)60-70℃烘箱中烘3-4小(xiao)時)。
在實驗室檢測煤質設備中,量熱儀有不可取代的作用,為確保您的量熱儀實驗測試工作的(de)正常進行請按以下步驟進行操(cao)作:
1. 開機(ji):打開機(ji)器電(dian)源使機(ji)器進入工作狀態;
2. 打開(kai)氧(yang)(yang)彈清理氧(yang)(yang)彈杯、干鍋、點火柱(zhu)等部(bu)件,然后在氧(yang)(yang)彈杯中(zhong)加入(ru)10mmL室溫蒸餾水
3. 把(ba)氧彈(dan)頭放在(zai)氧彈(dan)架上(shang),裝好點(dian)(dian)火絲,(點(dian)(dian)火絲要安裝牢(lao)固,保證與(yu)點(dian)(dian)火電極可靠的電接觸,且兩(liang)電極間的點(dian)(dian)火絲長(chang)度要一致)然(ran)后在(zai)點(dian)(dian)火絲中部系(xi)上(shang)點(dian)(dian)火棉線(xian);
4. 將稱量好的坩堝內的試樣(稱準到0.0001mg)放入點火電極下部的坩堝位,然后把點火棉線的一端插入煤樣,但不得將煤粉帶出。(如是熱容量標定的苯甲酸片則(ze)只需(xu)壓住(zhu)棉(mian)線即可);
5. 小心將氧彈螺帽擰緊(jin),此后氧彈要輕拿(na)輕放以(yi)免棉線脫落(luo)導致點火失敗;
6. 打開氧氣瓶閥門,調整減壓器輸出壓力為2.8-3MPa。再用不小于15秒的時間將氧彈充到此壓力后放入量熱儀內筒氧(yang)(yang)彈固定座上(當充氧(yang)(yang)壓(ya)力大于3.2MPa時應將(jiang)氧(yang)(yang)氣(qi)放(fang)掉調整壓(ya)力后從新充氧(yang)(yang));
7. 依據(ju)試樣為發(fa)熱(re)(re)量的測試或苯甲酸的系(xi)統熱(re)(re)容(rong)量標定選(xuan)擇進入發(fa)熱(re)(re)量數(shu)(shu)據(ju)輸(shu)入界(jie)面或熱(re)(re)容(rong)量數(shu)(shu)據(ju)輸(shu)入界(jie)面,將所需(xu)數(shu)(shu)據(ju)對應輸(shu)入后點實驗(yan)開(kai)始;
8. 實(shi)驗完成后(hou)(hou)取出氧彈,放(fang)氣(qi)后(hou)(hou)檢(jian)查氧彈內部有無(wu)試(shi)樣飛(fei)濺(jian),如有則(ze)本次(ci)(ci)試(shi)驗作廢,最(zui)后(hou)(hou)結(jie)果數據由屏幕和(he)打(da)印機給出。如需再次(ci)(ci)實(shi)驗則(ze)按2、3、4、5、7步驟(zou)依次(ci)(ci)重(zhong)復進行。