解析量熱儀測試結果不準確的幾種因素
2018-02-27 來自: 鶴壁市科達儀(yi)器儀(yi)表有(you)限公司 瀏覽次(ci)數(shu):2735
量熱儀測定發熱量,結(jie)果不(bu)準確原(yuan)因大致有以(yi)下六種:
原因一:熱容量標定的(de)不準確
熱(re)(re)容(rong)量(liang)(liang)(liang)標(biao)定的(de)(de)不準(zhun)確常會帶來系統誤差(cha)。這(zhe)多半是由于使(shi)用(yong)的(de)(de)苯甲酸(suan)不合格或計算熱(re)(re)容(rong)量(liang)(liang)(liang)時(shi)忘(wang)記加硝酸(suan)形(xing)成熱(re)(re)。應該使(shi)用(yong)經計量(liang)(liang)(liang)部門(men)檢定合格的(de)(de)二級(ji)基準(zhun)量(liang)(liang)(liang)熱(re)(re)物--苯甲酸(suan)來標(biao)定儀器的(de)(de)熱(re)(re)容(rong)量(liang)(liang)(liang),并(bing)且保證(zheng)計算正確。另外(wai),發熱(re)(re)量(liang)(liang)(liang)測(ce)定時(shi)的(de)(de)內筒(tong)水量(liang)(liang)(liang)與(yu)熱(re)(re)容(rong)量(liang)(liang)(liang)標(biao)定時(shi)的(de)(de)不完全一(yi)致(zhi),也會使(shi)標(biao)得的(de)(de)熱(re)(re)容(rong)量(liang)(liang)(liang)不適用(yong)于發熱(re)(re)量(liang)(liang)(liang)測(ce)定試(shi)驗(yan)。標(biao)定完熱(re)(re)容(rong)量(liang)(liang)(liang)后(hou)應將內筒(tong)水的(de)(de)質(zhi)量(liang)(liang)(liang)(包括內筒(tong)本身)記下(xia)來,保證(zheng)在以后(hou)所有的(de)(de)試(shi)驗(yan)中內筒(tong)水量(liang)(liang)(liang)完全一(yi)致(zhi)。
原因二:內筒中的水量(liang)不能保(bao)持一致
內筒水(shui)的熱(re)(re)容量(liang)在整個(ge)儀器(qi)的熱(re)(re)容量(liang)中(zhong)占(zhan)相(xiang)當(dang)大的比重,由于水(shui)的比熱(re)(re)很(hen)大,內筒中(zhong)水(shui)的量(liang)若不準確將會使儀器(qi)的熱(re)(re)容量(liang)發生變化,從(cong)而導致發熱(re)(re)量(liang)的測定誤差。
原因(yin)三(san):攪拌器故(gu)障或攪拌速度不均勻
有時熱量計的攪拌器接線(xian)部(bu)分接觸不良(liang),有虛接現象,這樣會導(dao)致攪拌速度(du)時(shi)快時(shi)慢,時(shi)轉時(shi)停;有時(shi)攪拌槳上得不合適,被卡住而不能自由動作,會導(dao)致內筒水局部(bu)獲得的熱不能及(ji)時(shi)均勻(yun)地散出(chu),從而使測得的內筒溫度(du)變化為的溫度(du)變化。用這種溫升計算出(chu)的發熱量(liang)必然(ran)是錯誤的結(jie)果。當發現熱容量(liang)或發熱量(liang)重復測定結(jie)果忽(hu)高忽(hu)低,相差很遠時(shi),可從這方面查找原因(yin)。
原因(yin)四:使用貝克曼溫(wen)度計未進行(xing)或未能正確地進行(xing)毛細孔徑和平均分(fen)度值的修(xiu)正。
使用(yong)貝克曼溫度(du)(du)計測(ce)(ce)量內(nei)筒溫度(du)(du)變化,若(ruo)不(bu)能正確(que)的(de)(de)進行毛細孔徑值和平均分(fen)度(du)(du)值的(de)(de)修正,將會使測(ce)(ce)得的(de)(de)溫升不(bu)準(zhun),從而導(dao)致發熱量測(ce)(ce)定的(de)(de)誤差。
原因五:使用絕(jue)熱式熱量計時,熱量計的平衡點有問題(ti)
若絕(jue)熱(re)(re)式熱(re)(re)量計(ji)的(de)平衡(heng)點(dian)(dian)(dian)(dian)有(you)(you)問題,將會(hui)(hui)影(ying)響外筒(tong)(tong)(tong)溫度(du)跟蹤內(nei)筒(tong)(tong)(tong)溫度(du)的(de)能力(li),使外筒(tong)(tong)(tong)溫度(du)常會(hui)(hui)與內(nei)筒(tong)(tong)(tong)不(bu)(bu)(bu)一致,導致內(nei)外筒(tong)(tong)(tong)之(zhi)間過多(duo)的(de)熱(re)(re)交(jiao)(jiao)換。而(er)在(zai)絕(jue)熱(re)(re)式熱(re)(re)量計(ji)中不(bu)(bu)(bu)對(dui)這種(zhong)熱(re)(re)交(jiao)(jiao)換進(jin)行(xing)校正(zheng),因此,這部分(fen)熱(re)(re)交(jiao)(jiao)換將會(hui)(hui)引起發(fa)熱(re)(re)量的(de)測(ce)定(ding)(ding)誤差。平衡(heng)點(dian)(dian)(dian)(dian)有(you)(you)問題,是(shi)指(zhi)平衡(heng)點(dian)(dian)(dian)(dian)不(bu)(bu)(bu)穩(wen)定(ding)(ding)或(huo)(huo)不(bu)(bu)(bu)出(chu)現(xian)。其原(yuan)因,一則(ze)是(shi)可能是(shi)操作者在(zai)使用熱(re)(re)量計(ji)前未(wei)仔細(xi)檢查和調(diao)定(ding)(ding)平衡(heng)點(dian)(dian)(dian)(dian),二則(ze)可能是(shi)儀器的(de)自動控溫線路有(you)(you)問題。平衡(heng)點(dian)(dian)(dian)(dian)不(bu)(bu)(bu)出(chu)現(xian)或(huo)(huo)不(bu)(bu)(bu)穩(wen)定(ding)(ding)的(de)表現(xian),是(shi)在(zai)較長時間內(nei)內(nei)筒(tong)(tong)(tong)溫度(du)不(bu)(bu)(bu)能達(da)到(dao)恒(heng)定(ding)(ding),或(huo)(huo)是(shi)這次(ci)平衡(heng)出(chu)現(xian)了,下次(ci)又不(bu)(bu)(bu)出(chu)現(xian)了。遇到(dao)這種(zhong)情況,操作者應(ying)該仔細(xi)檢查和調(diao)定(ding)(ding)儀器的(de)平衡(heng)點(dian)(dian)(dian)(dian),若經多(duo)次(ci)調(diao)試,總是(shi)不(bu)(bu)(bu)能成功,則(ze)應(ying)請廠(chang)家修理或(huo)(huo)幫助查找原(yuan)因。
原因六:自動(dong)熱(re)(re)量(liang)(liang)計的性能(neng)不十分穩(wen)定。 某些(xie)自動(dong)熱(re)(re)量(liang)(liang)計,特(te)別是偏離(li)經典量(liang)(liang)熱(re)(re)理論(lun)和量(liang)(liang)熱(re)(re)方式(shi)的自動(dong)熱(re)(re)量(liang)(liang)計,由(you)于設(she)計考(kao)慮不周或(huo)使用(yong)的傳感器(qi)等電(dian)子元器(qi)件質(zhi)量(liang)(liang)不穩(wen)定,使熱(re)(re)量(liang)(liang)計的測溫或(huo)其(qi)他性能(neng)發生(sheng)很大變動(dong),造成發熱(re)(re)量(liang)(liang)測定結果不準。
鶴壁科達儀器——煤質分析儀器廠家,生產各種型號量熱儀:全自動量熱儀、微機全自動量熱儀、雙控自動量熱儀、制冷恒溫全自動量熱儀,磚廠專用量熱儀等。另外還生產供應工業分析儀系列、定硫儀系列、膠質層測定儀系列、測氫儀系列、馬弗爐系列、灰熔點系列、機械化(hua)采、制樣設(she)備(bei)等。擁有完整(zheng)的(de)精密(mi)電子儀器、生產(chan)加工(gong)設(she)備(bei)和試(shi)驗手(shou)段。產(chan)品廣泛應(ying)用與煤炭、電力、化(hua)工(gong)、冶(ye)金、地質、大專院校及技術監督(du)等部門(men)的(de)實驗室、化(hua)驗室。提(ti)供上門(men)安裝調試(shi)培訓服務,如(ru)果您有任何(he)疑問,歡迎(ying)隨時(shi)聯系我們。